You are currently on the South Georgia and the South Sandwich Islands version.

Looks like you are in United States.

Switch to local content?

Technical Specifications

Heidelberg OPERA

Display Options for Surgeons
  • Digital binoculars
  • 55” 3D 4K-UHD monitor
  • 32” 3D 4K-UHD monitor (mounted to the microscope stand)
Display Options for Observers
  • Up to 4 via wireless transmitter (2D/3D, up to 4K-UHD resolution)
  • 1x 3D via HD-SDI
  • 2x 2D via HD-SDI
Camera Sensor Resolution2x 4K-UHD cameras (3840 x 2160 pixels)
MagnificationMotorized optical 6x zoom
50% digital zoom
Focus Range90 mm (in 0° position)
Working Distance200 mm
ApertureMotorized, 9 positions
Red Reflex IlluminationLED
Surrounding Field IlluminationLED
Built-in KeratoscopeLEDs
Surgical AnnotationsYes
Recording
  • 4K-UHD (3840x2160), 60fps
  • Full-HD (1920x1080), 60 fps
  • Full-HD (1920x1080), 30fps
  • 2D and 3D
Digital Filters
  • Color filters
  • Contrast enhancement
  • Sharpness
  • Digital brightness
Physical Filters
  • Retina protection filter (built-in)
  • Retina protection plate (built-in)
  • 532 nm laser filter (built-in)
Compatible Fundus Viewing Systems
  • Oculus BIOM® 6
  • Oculus BIOM® 5mI/cI
  • Oculus BIOM® ready/ready+
Interfaces
  • 1x HDMI optical output
  • 2x HD-SDI in
  • 2x HD-SDI out
  • Wireless video transmitter
  • USB 3.2 gen 2

Intraoperative OCT

Imaging method

Swept-source optical coherence tomography (SS-OCT) 

Center wavelength

1,060 ± 10 nm 

Scan frequency70 kHz
B-scan and cross scanB-scan and cross scan
Optical power at the cornea< 8.6 mW
A-scan imaging depth in air≥ 20 mm
Axial resolution in tissue≤ 7.5 µm*
Transverse resolution for anterior segment imaging≤ 30 µm
Anterior segment scan length≤ 20 mm
Number of A-scans per B-scan≤ 1,024
Scan rotation360°, adjustable in 0.1° increments
Conformity

ISO 16971-1:2024

Laser classification

Class 1 in accordance with IEC 60825-1:2014 

* Calculated using a tissue refractive index of 1.33