Switch to local content?

Go beyond the microscope

Heidelberg OPERA, with next-generation swept-source intraoperative OCT, seamlessly integrates real-time cross-sectional imaging into the surgical workflow.

A new era in digital surgery

Heidelberg OPERA with iOCT ushers in a new era of image-guided surgery by revealing anatomical structures beneath the tissue surface that cannot be seen through microscopy alone.

Intraoperative OCT

Bringing diagnostic quality OCT imaging into the OR

Extend the surgeon's view beyond the surface

Integrated iOCT extends the view beyond the visible surface, helping surgeons to assess tissue configuration, confirm anatomical relationships, and guide delicate surgical maneuvres in real time. As digital surgery continues to evolve, Heidelberg OPERA provides a platform for future innovations such as image-guided workflows, quantitative intraoperative measurements, and intelligent surgical assistance.

iOCT Technical Specifications

Imaging method

Swept-source optical coherence tomography (SS-OCT) 

Center wavelength

1,060 ± 10 nm 

Scan frequency70 kHz
B-scan and cross scanB-scan and cross scan
Optical power at the cornea< 8.6 mW
A-scan imaging depth in air≥ 20 mm
Axial resolution in tissue≤ 7.5 µm*
Transverse resolution for anterior segment imaging≤ 30 µm
Anterior segment scan length≤ 20 mm
Number of A-scans per B-scan≤ 1,024
Scan rotation360°, adjustable in 0.1° increments
Conformity

ISO 16971-1:2024

Laser classification

Class 1 in accordance with IEC 60825-1:2014 

* Calculated using a tissue refractive index of 1.33