HRT SpecificationsTable Footprint
|
|
|---|---|
| Instrument type: | Confocal scanning laser ophthalmoscope |
| Field of View: | 15° x 15° (transverse) |
| Scan Depth: | 1.0 to 4.0 mm (automatic |
| Optical Resolution: | 10 µm /pixel (transverse |
| Repeatability: | 20 µm |
| Digital Image Size: | 2D image: 384 x 384 pixels 3D image: up to 384 x 384 x 64 pixels |
| Image file size: | 30 MB uncompressed, 5 MB compressed (typical) |
| Scan Time: | 2D image: 24 milliseconds 3D image: 1 second typical (2mm scan depth) |
| Focus Range: | -12 to +12 diopters spherical -6 to +6 diopters cylindrical |
| Minimum Pupil Diameter: | ≥ 1 mm |
| Light Source: | 670 nm diode laser |
| Image alignment/ artifact rejection: |
TruTrack™ proprietary software |
| Display modes: | Single frame, Multiframe (movie), 2-D mapping, 3-D mapping |
| Portability: | Optional laptop version and optional carrying case |
| Software analysis: | Glaucoma, retina, cornea |
| Operating systems: | Heidelberg Eye Explorer and Windows XP |
| Networking: | Heidelberg Eye Explorer, IHE compliant, DICOM compliant (please read compliance statement) |
| Electronic Medical Record Partners: |
OfficeMate, Ifa, NextGen, MedFlow(call if your supplier is not listed |
| Glaucoma Module: | Moorfield’s Regression Analysis RNFL Analysis C/D Ratio Cup Shape Analysis Glaucoma Probability Score Ethic specific databases OU Symmetry analysis 9-point Quality Check Optic disc size adjusted On-line operator feedback Topographic Change Analysis (progression) |
| Retina Module: | Edema Index Retinal Thickness 3-D retinal thickness display ETDRS moveable grid Trend Analysis |
| Cornea Module: | 400×400 micron field of view 2 micron intervals, up to 80 sections 1 micron resolution Semi-automatic cell countion Section scan, sequence scan, volume scan |
- Speaker Presentation Videos Hot off the AAO Floor!
- New Spectralis® 3.2 Software Available at No Charge
- Spectralis HRA+OCT Found Helpful in Managing Geographic Atrophy
- Spectralis HRA+OCT possesses unique eye tracking and registration technology
- New research shows confocal microscopy useful for assessment of atopic keratoconjunctivitis
Heidelberg Engineering Academy
Contact Us
(800) 931-2230